(4) [MOD]
Contents:
Format:
Parameter:
Description & others:
Return value format:
a) Mode 00 return value format
Mode 00 repeatedly outputs the indicated value and status in order.
D1±xxxxxx<CR><LF>
Sabcdefgh<CR><LF>
• The return value ±xxxxxx is an indicated value with a decimal point.
If the indicated value has no decimal point, it is filled with '0' from the head.
• 'abcdefgh' describes status with '1' or '0.'
Numeric '1' is outputted to describe the following state.
a: Comparator H is [ON].
b: Comparator L is [ON].
c: Spare
d: Spare
e: During measurement
f: Detecting peak
g: During hold
h: Various errors occurred
• To stop the output, set the mode to Mode 10.
b) Mode 02 return value format
Mode 02 outputs the indicated value and status only once at hold.
D1±xxxxxx<CR><LF>
Sabcdefgh<CR><LF>
• Contents of the indicated value and status are as same as those of the Mode 00.
• Assume that it is operated in the Mode 00. If the Mode 02 is specified, the following message is sent
back.
ng<CR><LF>
c) Mode 10 return value format
Mode 10 operates the WGA-670B according to contents of the control commands.
ok<CR><LF>
Sets transmission mode
MODpp<CR>
pp (Transmission mode)
• 00: Repeatedly outputs the indicated value and status.
• 02: Outputs the indicated value and status only once at hold
mode.
• 10: Operation by control command.
This command becomes disabled with the power OFF. Power ON, and it is in
the Mode 10 operation mode.
38