Function
Selecting Mode Function Name
Initialize
F-12
Key Operation & WGA-670B
Movement in Function Setting Mode
•Key code enable setting
'00000' is indicated.
Here, set the indication to read '00628' or '11111'
and press the SET key. The following set values
return to the factory default state. Then, the key
code is enabled. Therefore, when returning the set
value to the factory default the next time, the
above operation is required.
To disable the key code, conduct the function
'F-12.'
•Key code disable setting
'00000' is indicated. Press the SET key as it is.
Then, the following set value returns to the
factory default state.
The key code is disabled. The set value returns to
the factory default stage even if no key code is
entered.
●Factory default state
Hold mode
Delay time
Specified time
Analog output
Smoothing
Additional value
Auto ZERO Compensation range
Determination time
ZERO near ZERO
Hysteresis
BCD logic
HIGH/LOW limit set value
HIGH limit
LOW limit
●Calibration value and Bridge Exciter value
(BV) are not initialized.
●The indication '00628' or '11111' is a key code
to prevent calibration value from carelessly
changing due to malfunction.
17
0-0-0
0 second
0 second
0 V with 0000 display
10 V with 4000 display
0-0-0
0
0
0.0 second
0
0
0-0-0
+19999
-19999
Functional
Descriptio
n
• A
function
that
returns
the set
values to
factory
default
state.