KYOWA WGA-670B Instruction Manual page 19

Instrumentation amplifier
Hide thumbs Also See for WGA-670B:
Table of Contents

Advertisement

Function
Function Name
Selecting Mode
F-04
Hold Mode
Self-Check
F-05
F-06
Analog Output
(D/A Output)
Scale
Key Operation & WGA-670B
Movement in Function Setting Mode
Indicates the following items (1), (2), and
(3).
(1) Setting Hold mode
0 : Tracking
1 : Arbitrary point hold
2 : Peak hold
3 : Block specified peak hold
4 : Time specified peak hold
(2) Display mode
0 : Normal display mode
1 : Hold display mode
(3) Comparison mode
0 : Normal comparison mode
1 : Hold comparison mode
Select and change numerics in items (1),
(2), and (3) and press the SET key.
Previously set delay time is indicated.
After it is changed, press the SET key.
Previously set specified time is indicated.
After it is changed, press the SET key.
●Delay time range
0.00 to 9.99 (sec)
●Specified time range
0.01 to 99.99 (sec)
'. . . . . ' is indicated.
Pressing the SET key at this time lights
up all the indicators to conduct checking.
The checked results are indicated as
shown in the right.
●If any faulty results appear on the
indicator, contact KYOWA or our
representatives.
Set the indication to set the analog output
value to 0V or 4mA and press the SET
key.
Next, set the indication to set the analog
output value (D/A output) to 10 V or 20
mA and press the SET key. Then, analog
output is scaled.
●Setting range: ±19999
14
Functional
Description
•Sets various Hold
functions.
•For details, see
"4-2-5."
For setting hold
display mode and
hold comparison
mode, hold modes 3
and 4 are enabled. 
 
Set the delay time
and then, the hold
operation start can be
delayed by the preset
time.
•A function that
checks the stored
memory.
Good: Normal
E-01: RAM failure
E-02: Flash memory
failure
•A function that
defines and sets
ZERO and full
scale values.