(5) [/M]
Contents:
Format:
Description & others:
(6) [GDT]
Contents:
Format:
Parameter:
Return Value Format:
Description & others:
(7) [GST]
Contents:
Format:
Parameter:
Return Value Format:
Description & others:
Writes the settings to non-volatile memory.
********/M<CR>
Add '/M<CR>' at the end of the determined command. Then, it is stored in
the non-volatile memory. If the power is turned OFF, the contents determined
without this command shall return to the original setting.
Outputs indicated value.
GDT1<CR>
None
D1±xxxxxx<CR><LF>
Return value ±xxxxxx is an indicated value including a decimal point. If the
indicated value has no decimal point, it is filled with '0' from the head.
Outputs the status.
GST<CR>
None
Sabcdefgh<CR><LF>
Return value abcdefgh represents the status with '1' or '0.'
Numeric '1' is output in the following state.
a: Comparator H is [ON].
b: Comparator L is [ON].
c: Spare
d: Spare
e: During measurement
f: Detecting peak
g: During hold
h: Various errors occurred
Corresponding command