Preventative maintenance procedures
2. Click on Start Guided Alignment to be guided through the
verification and (optional) alignment process. To skip guided
alignment, select the Manual Adjustment tab to adjust the dispense
location.
A cassette and plate are loaded in the first step. Any type of cassette can
Load and
be used for alignment. Place a clear lid or plate seal on the plate.
validate
After the cassette and plate are loaded, click Next and they will be
validated. If necessary, you will be prompted again for cassette and/or
plate load.
Thermo Fisher Scientific
Thermo Scientific™ Multidrop™ Pico 1 Digital Dispenser Operating manual
31