System
Errors
Detection
Problem
while in-
session
mode
2
("Detecting"
message will
appear in
the middle
of the circle).
INVU Sensor
Band™
3
Connectivity
Problem
Network
4
Connectivity
Problem.
42
Cause
There was a problem
with the detection of
the maternal heart
rate for more than
1 minute during the
monitoring session.
The INVU Sensor
Band™ is not properly
attached to the
abdomen.
Data transfer problem
between the mobile
device and the INVU
Sensor Band™ during
the session.
Wi-Fi network or
cellular Internet
provider is
disconnected during
the signal quality
phase.
Required Action
Re-attach the INVU Sensor
Band™ to your abdomen as
instructed in section 5.1
1.
Verify the INVU Sensor
Band™ and mobile device
are close and within
Bluetooth range: less
than 16 feet (5 meters)
away, and with no walls
or obstacles in-between.
2.
Disconnect other devices
connected to your Wi-Fi
or Bluetooth network.
1.
Make sure your Internet
connection is working
properly (Wi-Fi or cellular
data plan).
2.
Press CLOSE or
CONTINUE.
3.
If not successful,
consider switching
between Wi-Fi and
cellular data plan.