Code
Descriptioin
AL28
Maintenance of pump
 2
AL29
Maintenance of fan motor
AL30
Maintenance of comprssor
Contact
input1
AL31
detection
Contact
input
AL32
detection
Water leakage
AL33
Electric
AL34
resistivity/conductivity rise
Electric
AL35
resistivity/conductivity drop
Electric
resistivity/conductivity
AL36
sensor error
1: ''Stop" or "Continued" are default settings. The user can change them to "Continued" and "Stop".
2: For air-cooled type.
HRS Series
Table 3-2 Error Type and Alarm Set Value (2/2)
Operation
Continued
 2
Continued
Continued
signal
 1
Stop
2
signal
 1
Stop
Continued
Continued
Continued
Chapter 3
Alarm Indication and Troubleshooting
Cause / Remedy
(Press the reset key after eliminating the cause.)
The timing of a periodical check is informed.
Recommended to ask for the check and service of the
pump, fan motor and compressor.
Contact input is detected.
・Check if the leakage sensor is connected.
・Leakage occurred. Check the leakage point.
Electrical resistivity/conductivity is larger than the set
value.
If an electric conductivity sensor is used, replace the DI
filter.
Electrical resistivity/conductivity is smaller than the set
value.
If an electric resistivity sensor is used, replace the DI filter.
・ Check if the electric resistivity/conductivity sensor is
connected.
・ There may be short circuit or open wire of the
resistivity/conductivity sensor. Replace the sensor.