7. Options
24
7. Options
The operating interface makes it easy to configure optional equipment, and some of it can
be configured using the PC programme MAP Mon for Windows.
On MAP Check 900X-2 it is only possible to configure optional equipment by means of the
PC programme.
7.1. Automatic three-channel scanner
MAP Check 9000 can be equipped with an automatic three-channel scanner (multiplexer).
The scanner can be set to automatic switching between one, two or three measuring points.
When a scanner has been installed (indicated by means of parameter
menu (see "Setting up various extra equipment" on page 18), the
7.1.1.
SCAN
ti.1 ti.2 ti.3
•
These three parameters are used to determine the duration of measurement on channels
1, 2 and 3, respectively, in seconds. The entered value can range from 0 to 9999 seconds.
It is not possible to set all values to zero. If you try to do so,
seconds. When you switch to a new channel, flushing will always be carried out for del.
seconds.
prio
•
: (Channel 1 priority)
Giving a higher priority to channel no. 1. This parameter is used to determine the channel
selection sequence.
On
: The channels are selected in the following order: 1,2,1,3,1,2,1,3 ...
Off
: The channels are selected in the following order: 1,2,3,1,2,3,1,2 ...
The reason for giving channel no. 1 a higher priority is that when an MAP Con 9000
(proportional flow controller) is connected, adjustment is only carried out when channel no.
1 has been selected.
PA.SC
•
: (Scanner parameter)
This parameter consists of four digits, each of which can be set to '0' or '1'. It is used to
determine which of the three channels of the scanner must be monitored by the oxygen
alarms (
have a hold function.
PBI-220047-F 01/2008
menu
:(Timer (clock register) 1, 2, 3)
On
off
or
Al.1
Al.2
and
). The parameter is also used to determine whether the alarms must
Channel 1
alarm
0 = On
1 = Off
Digit
1
SET
DIA
cal
scan
reg.f
ti.1
ti.2
ti.3
prio
pa.sc
can be selected.
Channel 2
Channel 3
alarm
alarm
0 = On
0 = On
1 = Off
1 = Off
2
3
COPYRIGHT ©
MAP Check 9000 - User Manual
PA.3
in the setup
SCAN
menu is added:
ti.1
is automatically set to 10
Alarm hold
function
0 = Off
1 = On
4
PBI-Dansensor A/S
Need help?
Do you have a question about the MAP Check 9000 and is the answer not in the manual?