Filter Mat In The Power Supply Unit; Replacement Interval; Location; Removing The Cover - Dräger Medical Zeus Technical Documentation Manual

Anesthetic workstation
Table of Contents

Advertisement

Zeus
3
Filter mat in the
power supply unit
3.1
Replacement inter-
val
3.2

Location

3.3

Removing the cover

3.3.1
Plastic cover underneath
the pneumatic interface
3.4
Replacing the filter
mat
5133.001
For replacement intervals of the filter mat, refer to the Zeus Instructions for
Use manual.
The power supply unit is located on the side of the Zeus which holds the
breathing system
(Figure
Figure 4
Fitting location of the power supply unit in the Zeus
CAUTION
Risk of short circuiting during disassembly.
In order to avoid short-circuits in the device during disassembly of parts or
assemblies, shut down the Zeus, set the power switch to the "OFF" posi-
tion, and disconnect the Zeus from the power and gas supplies.
1. Remove the plastic cover underneath the pneumatic interface of the
breathing system as described in the Repair Instructions"Opening the
device", section
1.2.1 Plastic cover underneath the pneumatic
The filter mat is located behind a sheet-metal frame, which must be removed
in order to replace the filter mat.
Maintenance procedures
4, dashed frame).
interface.
83