Function description
9.3

Output voltages

58
The power supply unit delivers seven no-load and short-circuit-proof output
voltages (see also
Figure
UBlower, 48 V/1.5 A
UMain, 24 V/15 A
UHERMES, 24 V/3.5 A
ULED, 24 V/power limit 6...10 VA at 20...29 V
UPermanent, 24 V/0.03 A
UGA,12 V/3.5 A
VCC-ISO, 5 V/0.1 A
Figure 34 Power supply unit front view
Table 6
Legend to
Item
X02
Hermes computer
X03
Apollo module
X04
Transfer PCB (Power)
X05
Transfer PCB (Signals)
X06
Emergency-off button
X07
On button (Prod./Service)
X08
DAGMAR PCB
X09
Hermes monitor
X10
DIVA
X12
ZEUS motherboard
X13
Blower capacitor
F1
Apollo module (12 V/20 W), 3 A fast
F2
DIANA (24 V/12 W), 3 A fast