Batteries Of The Uninterruptible Power Supply; Replacement Interval; Location; Replacing The Rechargeable Batteries - Dräger Medical Zeus Technical Documentation Manual

Anesthetic workstation
Table of Contents

Advertisement

Zeus
11
Batteries of the unin-
terruptible power
supply
11.1
Replacement inter-
val
11.2

Location

11.3
Replacing the
rechargeable batter-
ies
5133.001
NOTE
Batteries represent special waste. Batteries must be disposed of in confor-
mity with local waste disposal regulations.
For replacement intervals of the Zeus rechargeable batteries, refer to the
Instructions for Use manual.
The rechargeable batteries for the uninterruptible power supply are located
underneath the trolley in a drawer. Before replacing the rechargeable batter-
ies, check if the fuse
(Figure
Figure 24 Fuse (1) and connector (2) on drawer unit
CAUTION
Risk of short circuiting during disassembly/assembly.
In order to avoid short-circuits in the device during disassembly of parts or
assemblies, shut down the Zeus, set the power switch to the "OFF" posi-
tion, and disconnect the Zeus from the power and gas supplies.
1. On the rear of the Zeus underneath the trolley, detach the PE conductor
and the connector
(Figure
2. Unscrew the hexagon socket screws in the trolley foot in the area of the
blue drawer (1 each at the front and rear).
3. Hold the cables out slightly and withdraw the drawer for the rechargeable
batteries.
CAUTION
Risk of short circuiting during disassembly/assembly.
To avoid shorting when detaching/attaching the cables, disconnect the
cable connecting the two rechargeable batteries first upon disassembly
and reconnect it last upon reassembly.
Maintenance procedures
24/1) is OK.
24/2) from the rechargeable batteries' drawer.
105