Zeus
6
Zeus Gas Measur-
ing Module (GMZ)
6.1
General introduction
6.1.1
Purpose
6.1.2
Layout
5133.001
Figure 19 GMZ
The function of the Zeus gas measuring module (GMZ) is to measure the
gases O2, N20 and CO2 and the anaesthetics Desflurane, Enflurane, Hal-
othane, Isoflurane and Sevoflurane for the purposes of monitoring. Two inde-
pendent systems are used for this. One is a closed-loop patient-local system
which taps the measured values directly at the Y-piece. The second system
taps the measured values in the inspiratory branch. It monitors the loop con-
trol function and shuts down the loop in the event of an error.
The GMZ comprises the component assembly:
–
Patient gas analyser
–
IRIA
or
ILCA 2
–
Servomex
or
Pato
–
Dräger water trap
Function Description
(PGA) with
39