Cover Violation - Agilent Technologies Infinity 1260 User Manual

Refractive index detector
Hide thumbs Also See for Infinity 1260:
Table of Contents

Advertisement

7
Error Information
General Error Messages

Cover Violation

The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the
foam is removed while the lamps are on (or if an attempt is made to switch on
for example the lamps with the foam removed), the lamps are switched off,
and the error message is generated.
Probable cause
1
2
106
The top foam was removed during operation. Please contact your Agilent service
Foam not activating the sensor.
Suggested actions
representative.
Please contact your Agilent service
representative.
Agilent 1260 Infinity RID User Manual