GLOBAL COMMANDS
The following global commands affect all power supplies on a multi-drop chain. See
section 9.4.2 for requirements. There is no query version of these commands
9.4.4.
Global Set the Voltage Limit
This is the global version of the :VOLT command in section 7.2.1
Syntax:
GLOBal:VOLTage[:LEVel][:IMMediate][:AMPLitude] <nn.nn>
Parameter:
<nn.nn> is volts, the setting applied to all power supplies
Example:
GLOB:VOLT 9.45
9.4.5.
Global Set the Current Limit
This is the global version of the :CURR command in section 7.2.2
Syntax:
GLOBal:CURRent[:LEVel][:IMMediate][:AMPLitude] <nn.nn>
Parameter:
<nn.nn> is amps, the setting applied to all power supplies
Example:
GLOB:CURR 350
9.4.6.
Global Set the Output ON or OFF
This is the global version of the OUTP:STAT command in section 7.2.3
Syntax:
GLOBal:OUTPut:STATe <0|1|OFF|ON>
Example:
GLOB:OUTP:STAT 1 (or GLOB:OUTP:STAT ON)
9.4.7.
Global Reset Power Supplies
This is the global version of the *RST command in section 7.7.2
Syntax:
GLOBal:*RST
Example:
GLOB:*RST
9.4.8.
Global Save All Settings
This is the global version of the *SAV 0 command in section 7.7.3
Syntax:
GLOBal:*SAV 0
Example:
GLOB:*SAV 0
9.4.9.
Global Recall All Settings
This is the global version of the *RCL 0 command in section 7.7.4
Syntax:
GLOBal:*RCL 0
Example:
GLOB:*RCL 0
9.4.10. Error and Status Registers
The error and status registers described in sections 7.8.5 through 7.8.17 operate somewhat
differently in a multi-drop system.