10.6.5 Adjusting sterile water volume
10.6.6 Removing sterile water bag holder (OP deliveries)
User's manual
Use the clamp on the sterile water tube to adjust the water volume. Push the
slide forward to reduce the volume, and backward to increase the volume.
The sterile water bag holder that is mounted on the column can be removed
from its mount as follows:
1. Loosen the screw on the mount with a 4 mm Allen key.
2. Remove the sterile water bag holder.
3. Tighten the screw.