Setting Up The Apparatus (Practical Exercise) - Datcon DT9502 Operating Instructions Manual

Intrinsically safe temperature meter / transmitter
Table of Contents

Advertisement

DT9502
Task
Entering the menu
Setting the temperature
value that belongs to 4
mA
20171205-V3

10.7. Setting up the apparatus (practical exercise)

There is a measurement loop, in which the 4 mA current
should be assigned to 100°C, and 20 mA should be
assigned to 200°C.
Let's have the following settings for DT9502:
The temperature value belonging to the 4 mA current:
100.0 °C.
The physical value belonging to the 20 mA current: 200.0
°C.
The leader nulls with no value are not written on the
display.
Use the supervisor code to enter the menu.
Read chapter 7.1. Writing the code in to see how the code
should be written in. On the dipslay you see this:
1. Keep stepping with the ▲ button as long as you see this

menu item:
2. Enter this menu item by using the ENTER button.
3. By the ▲ button you may increase the value of the
blinking digit.
4. By the ► button you may step to the next digit.

5. Set the value:
6. Press the MENU button. With this you leave the menu
item, and you see this:
.
.

.
