6.3 Prepare the instrument for measurement

1. Push Home to go to the main menu.
2. Connect the USB key to the USB port on the side of the instrument. The applications
on the USB key show on the display.
3. Push the arrow keys to highlight and select an application to install. Push the left or
right arrow key to select it. Do this step again to select additional applications to
install.
4. Push Import to install the selected applications.
5. Push OK to complete the installation. The installed applications show on the home
screen.
Note: To install more applications, push Home to go to the home screen, then remove the
USB key and reconnect it.
1. From the home screen, select Purge. All attached devices are listed.
2. Select All elements to purge all the attached devices, or select one device to purge.
Push Select. Air is removed from the device and filled with liquid from the bottle.
3. Push OK when the operation has completed.
4. Make sure that there are no air bubbles in the device. Do step
air bubbles.
5. Select the next device to purge if individual devices are being selected.
6. Push Exit when all the tubes are filled with reagent and the device has no air
bubbles.
Note: If a few small air bubbles can be seen on the inner wall and/or piston of the syringe, they
can be left without effecting system performance.
Startup
2
again if there are any
27