5.3 Prepare the instrument for measurement
1. From the main menu, select
listed.
2. Select All elements to purge all the attached devices, or select one device to purge.
Push Select. Air is removed from the device and filled with liquid from the bottle.
3. Push OK when the operation has completed.
4. Make sure that there are no air bubbles in the device. Do step
air bubbles.
5. Select the next device to purge if individual devices are being selected.
6. Push Exit when all the tubes are filled with reagent and the device has no air
bubbles.
Note: If a few small air bubbles can be seen on the inner wall and/or piston of the syringe, they
can be left without effecting system performance.
Purge, then push Start. All attached devices are