Hach TitraLab KF1000 Series User Manual page 62

Hide thumbs Also See for TitraLab KF1000 Series:
Table of Contents

Advertisement

Change application parameters
Parameter
Type
Part of name
Recommended electrode
Working mode
Current (sample analysis) Set the value of the imposed current during sample analysis
Current (titrant calibration) Set the value of the imposed current during titrant calibration
Parameter
Description
Name
Name of the titrant
Titrant concentration
Nominal concentration of the titrant
Resolution
Number of decimals of the nominal concentration of the titrant
Real concentration
Real concentration of the titrant used for result calculation
Resolution
Number of decimals of the real concentration of the titrant
Active
Select No to skip this method during titrant calibration
Calibration frequency Recommended frequency of titrant calibration (0 = no reminder)
Stirring speed
Speed of the stirrer during titrant calibration
Delay
Stirring time before the titration starts
Injection autodetect
Automatic start after the detection of the sample injection
End point
Ordinate of the end point
Proportional band
Proportional band of the PID regulation
Drift threshold
Drift value below which the cell is considered as dry
Max. addition speed
Maximum speed for titrant addition on titration regulation
Min. titr. duration
Set the minimum titration duration
Max. titr. duration
Set the maximum titration duration
Resolution
Maximum number of decimals to show for the real concentration of the
titrant
Titrant cal. limit check Activate the titrant calibration result boundary check
Min. titrant conc.
Minimum value of the real concentration of the titrant
Max. titrant conc.
Maximum value of the real concentration of the titrant
Standard name
Name of the standard
Standard amount
Amount of standard to be used for titer determination
Resolution
Maximum number of decimals for the amount value of the standard
Min. amount
Minimum amount of the standard
60
Table 21 Electrode
Description
Type of measurement electrode
Part of the name used to identify matching electrodes if the
preferred electrode is not connected
Recommended electrode to be used for this analysis
Specify if the probe is used in imposed voltage/current (AC or DC)
Table 22 Titrant
Titrant calibration
Incremental addition
Titration calculation
Basic mode Expert mode
x
x
x
x
Basic mode Expert mode
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x

Advertisement

Table of Contents
loading

Table of Contents