Hach TitraLab KF1000 Series User Manual page 53

Hide thumbs Also See for TitraLab KF1000 Series:
Table of Contents

Advertisement

Error/Warning
Format of message
received from the balance
is not supported
No connection with balance
Temperature sensor not
detected
Incorrect input value
No operator recorded
Serialization error
USB mass storage writting
error
Table 14 Syringe verification
Description
The data received from the balance is not
compatible with the format of messages
accepted by the instrument (fully
automated mode only).
The instrument is not receiving data from
the balance (fully automated mode only).
The instrument is not receiving data form
the temperature sensor (fully and partially
automated modes only)
The input value is out of range (fully and
partially automated modes only).
The operator value is not defined on the
syringe verification screen.
There was an error during the results
export.
Error when exporting data to the USB
storage device
Possible solution
Configure the balance to be compatible with the
format of messages accepted by the instrument
(refer to
Install a balance
on page 19)
Make sure that the weight measurement is stable
and did not get to the maximum weight.
Make sure that the balance cable is installed
correctly.
Make sure that the temperature sensor is installed
and connected correctly.
Make sure that the syringe verification
configuration values are correct.
Set an operator on the syringe verification
configuration.
Remove the USB storage device and make sure
the USB icon is no longer seen in the header bar,
then install the device again try the export again.
Try again, if the problem continues, contact
technical support.
Make sure that the USB storage device is
correctly installed and the USB icon can be seen
in the header bar (refer to
Troubleshooting
Table 3
on page 22).
51