Maintenance
To replace the pump cassette, refer to the illustrated steps that follow and obey the
instructions on the display.
8.7.5 Syringe verification
Use the syringe verification option to examine the syringe accuracy based on the ISO
8655 standard. For the best results, follow the procedure in the application note supplied
with the syringe verification kit.
Make sure to fully rinse the syringe with methanol dry before the test to remove titrant.
Make sure to fully rinse the syringe with methanol dry after the test to remove water.
Note: A license key is necessary to enable the syringe verification option on the instrument.
Contact the manufacturer or a sales representative to get a license key.
1. From the home screen, select Settings, then push Select.
2. From the settings screen, select Options, then push Select.
3. Push the left or right arrow key to select the expert mode, then push Done.
43