7.6 Purge

7.7 Cell management

Option
Description
Export data log Exports measurement data from the system to an external device. Preview
data selection before it is exported. Make sure that an external device is
connected to the instrument (e.g., a USB key, external hard drive, etc.).
Delete data log Removes measurement data from the system. Previews data selection
before it is removed.
3. Specify the data selection parameters. Push the left and right arrow keys to make a
selection. Push the up and down arrow keys to select an option.
Option
Result type
Application
Date
Operator
4. If View data log was the selected option, push View to see the selected data.
Push the up and down arrow keys to select a line of data and push Detail to see
more data
If only one Application is selected push the up and down arrow keys to select a
line of data and push Detail to see more data, or push the left and right arrow
keys to see related graphs
5. If Export data log or Delete data log was the selected option, push Preview to see
the selected data, then push Export or Delete to start the procedure.
Use this procedure to remove air bubbles from the system. Refer to
instrument for measurement
Use this option to fill or empty the measurement cell with the pumps. These procedures
stop automatically after a pre-determined time. The time remaining is shown against
Security time on the display.
1. From the home screen, select Cell management.
2. Push the left and right arrow keys to make a selection.
Option
Description
Fill cell (pump
Use this option to fill the measurement cell with reagent from pump 1.
1)
Empty cell
(pump 2)
Chemical exposure hazard. Before selecting this option, examine the
level of liquid in the waste bottle. Make sure that there is enough space in
the bottle for all of the liquid from the measurement cell.
Use this option to empty the measurement cell with pump 2.
Stop pumps
Use this option to stop the Fill cell and Empty cell procedures.
Stir
Use this option to start or stop the magnetic stirrer. This option can also be
used when the Fill cell and Empty cell procedures are operating.
3. Push Exit to go back to the home screen.
Description
Sets the type of result available.
Sets the available applications.
Sets the date range.
Sets the available operators.
on page 27 for instructions.
C A U T I O N
Standard operations
Prepare the
37