Maintenance

8.7.6 Other maintenance options

8.8 Storage and transportation

8.8.1 Prepare the instrument for short-term storage

44
4. From the home screen, select Maintenance, then push Select.
5. From the maintenance screen, select Syringe verification, then push Select.
Note: When the option is selected for the first time, enter the license key.
6. Push the up and down button to select and change the syringe test options.
Option
Description
Syringe
Selects the syringe. Used when there are two syringes.
Shows the syringe information: syringe number and volume.
Operator
Selects the operator ID
Configuration Sets the test settings: test volume and number of replicates. The total
dispensed liquid for the test shows when the two parameters are set.
Note: Make sure to use the applicable beaker and balance for the test based
on the amount of total dispensed liquid.
A field to write the test conditions is available in the test conditions settings.
Start
Starts the syringe verification procedure.
7. Push Start to start the syringe verification procedure. Follow the procedure supplied
with the syringe verification kit.
8. To cancel the syringe verification test and not save data to the data log, push Stop.
9. To cancel the syringe verification test and save the syringe verification results for fully
tested volumes to the data log, push Skip.
10. When the test is complete Pass or Fail shows. Push the left or right arrow to show the
results for the different volumes. Push the up and down arrow to show more results
data for the selected volume.
1. Select an option, then push Select.
Option
Stirring activation
Reagent
replacement
Maintenance
schedule
Maintenance
summary
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to the
current safety data sheets (MSDS/SDS) for safety protocols.
Use this procedure if storing the instrument overnight or over a weekend.
Description
Do a check of the magnetic stirrer. Push the up and down arrow keys
to increase or decrease the stirring speed.
Use this option to replace the reagents. Obey the instructions on the
display.
See the list of maintenance tasks. Push Edit to change the default
value.
See the number of days remaining for maintenance tasks. After doing
a task, push Reset to set the number of days remaining to the default
value.