Startup

6.1.1 Cell standby

6.2 Install the applications

26
Option
Description
Options
Sets the following options:
Note: When the instrument is set to on, the application parameters view is
always set to basic mode. Changes made to application data with the
instrument in expert mode are kept.
Security
Change the password. Specify which of these options are password
protected:
3. Push Back.
Make sure that the measurement cell is prepared before selecting an application,
because the cell standby starts automatically.
Cell standby dries the measurement cell automatically using a regulated addition of the
titrant as defined in the application. When the cell standby option is active:
Cell standby starts automatically as soon as an application is selected from the home
screen
The stirring speed is the speed specified for the application
Cell standby information is shown at the bottom of the application start screen
After the weight of the sample is validated at the end of the measurement procedure,
the cell conditioning automatically starts again
Cell standby continues until the next sample is selected and a new measurement
procedure starts
If the Cell management option is selected at the end of the measurement procedure,
cell standby is paused and starts again when the cell management procedure
completes
The measurement procedure starts when Start is selected. If the cell is dry, cell
conditioning is not needed and the user is asked to add the sample. If the cell is not
dry, cell conditioning starts
Use the supplied USB key to install the applications. The instrument can install a
maximum of five applications. The installed applications are shown on the top line of the
display.
If any errors occur during installation, refer to
Set the application parameters view to expert or basic mode
When the instrument is set to off, sets the syringe to empty into the
titrant bottle
Changes the temperature display unit
Set the cell standby parameter (refer to
Prints the measurement curve if a printer is connected
Specify if a balance is connected
Settings (all items except Info)
Maintenance schedule and summary
Delete data log
Export data log
Electrode calibration
Titrant calibration
Sample volume calibration
Syringe verification test
Troubleshooting
Cell standby
on page 26)
on page 47.