Section 8

8.1 Maintenance schedule

Task

Clean the instrument

on page 39. Also do this procedure immediately after any liquid spills on
the instrument.
Clean the sensor
on page 40.
Regenerate the molecular sieve
Examine all tube connections for leaks and damage. Replace as necessary. Refer to
Replacement parts and accessories
on page 40.
Replace the syringe. Refer to
Replace the Viton O-ring in the measurement cell. Refer to
on page 53 for part numbers.
Replace syringe electro-valve block. Refer to
on page 40.
Examine the syringe connections and inlet and outlet tubes for leaks and damage. Replace as
necessary. Refer to
Replacement parts and accessories
Replace the tubes
on page 40.
Replace the pump cassettes. Refer to
Examine the pump connections and inlet and outlet tubes for leaks and damage. Replace as
necessary. Refer to
Replacement parts and accessories
Replace the tubes
on page 40.
Examine the bottle caps and connections for leaks and damage. Replace as necessary. Refer
to
Replacement parts and accessories
8.2 Clean the instrument
Maintenance
Multiple hazards. Only qualified personnel must conduct the tasks described in this
section of the document.
Do not disassemble the instrument for maintenance. If the internal components must be cleaned
or repaired, contact the manufacturer.
Table 7
shows the recommended schedule of maintenance tasks. Facility requirements
and operating conditions may increase the frequency of some tasks.
Table 7 Maintenance schedule
on page 40
on page 53 for part numbers and to
Syringe management
on page 42.
Replace the syringe electro-valve block
Pump cassette replacement
on page 53 for part numbers.
Never use flammable or corrosive solvents to clean any part of the instrument. Use of these
solvents can degrade the environmental protection of the instrument and may void the warranty.
Clean the exterior surface with a moist cloth or with a mixture of water and mild detergent.
Dry with a soft cloth.
C A U T I O N
N O T I C E
Replace the tubes
Replacement parts and accessories
on page 53 for part numbers and to
on page 42.
on page 53 for part numbers and to
N O T I C E
1 day 7 days 12 months
x
x
x
x
x
x
x
x
x
x
x
39