Installation
Figure 3 Titrant tube and sensor positions
1 Titrant tube

3.8 Install the titrant and the reagent

16
2 Sensor
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to the
current safety data sheets (MSDS/SDS) for safety protocols.
Put a filled desiccant cartridge into the adapter on the titrant and solvent bottle caps.
Note: The molecular sieve in the filled cartridges must be regenerated regularly depending on the
humidity of the laboratory. Refer to
Loosen the tube connector on the bottle cap. Push the inlet tube through the connector.
Make sure that the end of the tube is at the bottom of the bottle. Tighten the connector on
the bottle cap.
Pump 1 is used to fill the measurement cell with solvent. Pump 2 is used to empty the
measurement cell. Refer to the illustrated steps that follow.
3 Magnetic stir bar
C A U T I O N
Regenerate the molecular sieve
on page 40.