The parameter setting of a connected pyrometer happens via "record data" sets. The I/O
module allows diagnostics message to be locked or released. Once all parameter settings are
performed, the I/O device signals that it is ready to send cyclic productive data.
Certain pyrometer characteristics are parameterizable during the configuration. The parameters
of the pyrometer substitute are used to set the overall settings of the PROFINET I/O node.
Some of the setting are used in the module as default settings and can be optionally overwritten
within the module configuration.
Temperature unit
Color mode
Slope
Emissivity
Transmissivity
Sensor offset
Sensor gain
Averaging time
Valley hold time
Peak hold time
Setpoint relay
Deadband
Decay rate
Relay alarm output control
Laser control
Panel lock
Analog output mode
Bottom temperature of output
Top temperature of output
Message diagnostics alarm
Message process alarm
Behavior on module fault

8.7.4. Structure of the input/output data

The input data length is 23 Byte.
1.800.561.8187
8.7.3.1. Pyrometer parameters
Parameter
8.7.3.2. Profinet alarm behavior
Parameter
8.7.4.1. Pyrometer module input data
www.
Description
Set the temperature unit
* 1000 (0.9  900)
* 1000 (0.9  900)
* 1000 (1.0  1000)
* 0.1s (1s  10)
* 0.1s (1s  10)
* 0.1s (1s  10)
in °C /°F
Set output mode
Set bottom temperature of
analog output
Set top temperature of analog
output
Description
The diagnostics information of
pyrometer is not transferred to
the PROFINET IO controller
is transferred to the PROFINET
IO controller
The process alarm of pyrometer
is not transferred to the
PROFINET IO controller
is transferred to the PROFINET
IO controller
.com
Options
PROFINET IO (Communication Option 1)
Setting
Celsius
Fahrenheit
1 , 2 color
850 ... 1150
100 ... 1100
100 ... 1100
-200 ... +200
800 ... 1200
0 ...3000
0 ...3000
0 ...3000
dev. range min. max
1 ...99
0 ...9999
normally open,
normally closed,
permanently open,
permanently closed.
off / on / flashing/ trigger
locked / unlocked
0 ... 20 mA / 4 ... 20 mA
0...9999°C /°F
0...9999°C /°F
Setting
message inactive
message active
message inactive
message active
set process data to zero,
set process data to last value