Fluke Endurance Series User Manual page 46

Innovative high temperature infrared pyrometers
Hide thumbs Also See for Endurance Series:
Table of Contents

Advertisement

Order of subentry appearance:
1. Subentry: RS485 BAUD RATE
2. Subentry: MULTIDROP ADDR.
3. Subentry: TERMIN. RESISTOR
4. Subentry: ETHERNET DHCP
5. Subentry: ETHERNET IP
6. Subentry: ETHERNET NM
7. Subentry: ETHERNET GW
8. Subentry: ETHERNET PORT
9. Subentry: WEB SERVER
1.800.561.8187
The subentry RS485 BAUD RATE is to set the RS485 communication baud rate,
whereat the default baud rate is set to 38400 bps
With the ▲ ▼ keys, you can toggle between the following communications baud rates:
1200 bps, 2400 bps, 9600 bps, 19200 bps, 38400 bps, 57600 bps, 115200 bps
The subentry MULTIDROP ADDR. is to assign a specific serial multidrop address to
an Endurance device, which is working in a 2-wire (half duplex) multidrop environment,
where several devices interact with each other.
With the ▲ ▼ keys, you can toggle between sub-addresses from 000 to 032
The subentry TERMIN. RESISTOR is to reduce signal reflections over long distance
connections by inserting a termination resistor of 120Ω.
With the ▲ ▼ keys, you can toggle between ON and OFF (120Ω insertion)
The subentry ETHERNET DHCP is to indicate to a network DHCP server, to obtain a
dynamic Ethernet address. The DHCP server assigns the Endurance device a dynamic
address out of an address pool.
With the ▲ ▼ keys, you can toggle between ON and OFF (dynamic address service)
The subentry ETHERNET IP is to set a fix unique network device address for the
Endurance device, if DHCP in inactive. The assigned address has to fit in the network
address pool of your subnet.
The ENTER button selects in a consecutive way the IP address byte aaa.bbb.ccc.ddd
With the ▲ ▼ keys, you can toggle the individual address byte between 0 and 255
The subentry ETHERNET NM is to set a fix unique network mask address to integrate
the Endurance device in an existing subnet domain. The assigned address has to fit in
the network address pool of your subnet.
The ENTER button selects in a consecutive way the NM address byte aaa.bbb.ccc.ddd
With the ▲ ▼ keys, you can toggle the individual address byte between 0 and 255
The subentry ETHERNET GW is to set a fix unique network gateway address to
integrate the Endurance device in an existing subnet domain. The assigned address
has to fit in the network address pool of your subnet.
The ENTER button selects in a consecutive way the GW address byte aaa.bbb.ccc.ddd
With the ▲ ▼ keys, you can toggle the individual address byte between 0 and 255
The subentry ETHERNET PORT is to set a fix port address for the relevant network
services of the Endurance device in an existing subnet domain. The assigned port
address is used for any special network request by the Endurance device.
With the ▲ ▼ keys, you can toggle the TCP/UDP port address from 0 to 65535
The subentry WEB SERVER is to activate the Endurance device internal web server
functionality for video and web based applications.
With the ▲ ▼ keys, you can toggle the WEB SERVER selector between OFF and ON
www.
The control panel menu structure and their associated entries
information@itm.com
.com
Device Control