Endurance® Series
Users Manual
Response Time at 95% Reading
E1ML
E1MH
E2ML
E2MM
E2MH
E3ML
E3MH
E1RL
E1RH
E2RL
Selectable Analog Current output
Emissivity Coefficient (1-Color mode)
E1ML, E1MH, E2ML, E2MM, E2MH,
E3ML, E3MH, E1RL, E1RH,
E2RL
Slope Coefficient (2-Color mode)
E1ML, E1MH, E2ML, E2MM, E2MH,
E3ML, E3MH
E1RL, E1RH, E2RL
Signal Processing
Averaging Range
Peak Hold Range
Valley Hold Range
Noise Equivalent Temperature (NET)

3.4. Optical Specifications

Optical Specifications
Optical Resolution D:S
E1ML:
E1MH:
E2ML:
E2MM:
E2MH:
E3ML:
E3MH:
E1RL (1C / 2C-mode):
E1RH (1C / 2C-mode):
E2RL (1C / 2C-mode):
Lens Options
Sighting Options
1.800.561.8187
12
2 ms
2 ms
2 ms
2 ms
2 ms
20 ms
20 ms
10 ms
10 ms
20 ms
0-20mA or 4-20mA (galvanic isolated)
16bit resolution, max. impedance: 500Ω
Digitally adjustable in increments of 0.001
0.100 to 1.100
0.100 to 1.100
0.100 to 1.100
Digitally adjustable in increments of 0.001
N/A
N/A
0.850 to 1.150
Averaging, Peak hold or Valley hold
0.1 to 299.9 s (300 s = )
0.1 to 299.9 s (300 s = )
0.1 to 299.9 s (300 s = )
1°C peak to peak, target emissivity of 1.00,
unobscured target, 3°C peak to peak for all
specified attenuation conditions
(assumes 95% energy at the focus point)
D:S = 160:1
D:S = 300:1
D:S = 160:1
D:S = 160:1
D:S = 300:1
D:S = 100:1
D:S = 300:1
D:S = 100:1
D:S = 150:1
D:S = 75:1
F0: 190 - 300mm
F1: 300 - 600mm
F2: 600mm - infinity (24" - ∞)
Visual/Laser
Visual/Camera
Visual/LED
www.
.com
(7.5 - 12")
(12 - 24")
information@itm.com