Excelon EZ Operator's Manual
70
Size Calibration
Procedures for size calibration
① Replace
holder with Standard JIG.
② Press the [F1] button to
start size calibration.
③ After calibration, press the
[F1] button to save or [F2]
button to discard.
Feeler Calibration
Same as above size calibration procedure except step ②.
Press the [F2] button to start feeler calibration.
Lens
adaptor