Image 3-7
CAUTION: When the display is assembled in the medical system, take care of the fixation of all
cables, to avoid unwanted detachment.
3.5 VESA mount installation
To install the display on a VESA mounting solution
The display can be attached to a VESA 100 mm or VESA 200 mm arm or stand.
Image 3-8
The VESA mounting holes at the back of the display are provided with M4-type blind fasteners to fix the VESA
mounting plate. Depending on the VESA plate thickness (T) and the thickness of possible washers (W), a
different screw length (L) should be selected.
Please respect the following rule to select an appropriate screw length:
•
L
= T + W + 15 mm
min
•
L
= T + W + 18 mm
max
Installation
R5911692 /01
MDSC-8231
21