Evita 2 dura – Cleaning, disinfection,
and sterilization
Disinfecting/Cleaning
Use surface disinfectants. For surface compatibility, use disinfectants based on:
— aldehydes,
— quaternary ammonium compounds.
To avoid the possibility of damage to material, do not use any disinfectants based on:
— alkylamine-based compounds
— phenol-based compounds,
— halogen-releasing compounds,
— strong organic acids,
— oxygen-releasing compounds.
Parts must not be sterilised in ethylene oxide – health hazard!
To avoid all risk of infection by hospital staff and other patients, the ventilator must be disinfected and cleaned before it is used on the next patient (Use protective clothing, eye protection, etc.).