Evita 2 dura – Operating concept
Ventilation parameters
Setting ventilation parameters
1. To set a ventilation parameter = press the corresponding
parameter key. The yellow LED in the key lights up
2. To set the value of the ventilation parameter = turn the
rotary knob. The value is displayed next to the parameter
key
3. To confirm the value = press the rotary knob
The yellow LED turns off
4. To prevent settings being changed inadvertently press the
Lock key. Its yellow LED will light up. The parameter keys
and ventilation mode keys will be protected against
inadvertent setting.
Before setting a new value press the Lock key again. The
yellow LED turns off.