Evita 2 dura – Cleaning, disinfection,
and sterilization
Disassembling
CO
sensor (optional): (1) Remove from the
2
cuvette. Unplug the connector from the back of the
unit.; (2) Remove the cuvette of the CO2 sensor
from the Y-piece.; (3) Remove the catheter cone
from the cuvette; Prepare the CO2 sensor for wipe
disinfecting; Reprocess the reusable cuvette e. g. in
a washer-disinfector (do not use rinse aid); The
disposable cuvette cannot be reprocessed and
must be disposed of.
© Drägerwerk AG & Co. KGaA, 2019
Temperature sensor: Remove from the
Y-piece or from the mounting of breathing
circuit; Do not pull the cable; Unplug the
connector from the back of the Evita 2 dura;
Prepare the temperature sensor for wipe
disinfecting; the temperature sensor is
designed for wipe-disinfection; the
temperature sensor must not be placed in a
cleaning and disinfection machine or
disinfectant bath.
Nebulizer (optional): (1) Remove the
nebuliser hose from the nebuliser and from
the port on the device; (2) Remove the
medication nebuliser from the adult
breathing circuit.
Nebulizer (optional): (1) Remove the
nebuliser hose from the nebuliser and from
the port on the device; (2) Dismantle the
medication nebuliser from the paediatric hose
system; (3) Remove the catheter connector
from the inlet; (4) Remove adapter from the
outlet. (5) Remove corrugated hose from the
adapter; Dismantle and reprocess the
medication nebuliser in accordance with its
specific Instructions for Use. Reprocess the
adapter parts in a washer-disinfector.