50 Specifications
Test specifications for enclosure port immunity to RF wireless communications equipment
1720
1845
1970
2450
5240
5500
5785
NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the
monitor may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a
For some services, only the uplink frequencies are included.
b
The carrier shall be modulated using a 50 percent duty cycle square wave signal.
c
As an alternative to FM modulation, 50 percent pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.
FCC compliance statement
This device complies with part 15 of the FCC Rules. Operation is subject to the following
two conditions:
This device may not cause harmful interference.
This device must accept any interference received, including interference that may
cause undesired operation.
Note
Vision Screener radio
The vision screener radio operates on 802.11 networks.
1700 - 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE Band
1, 3, 4, 25;
UMTS
2400 - 2570
Bluetooth,
WLAN, 802.11
b/g/n, RFID
2450,
LTE Band 7
5100 - 5800
WLAN 802.11
a/n
This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to Part 15 of FCC Rules. These limits are
designed to provide reasonable protection against harmful interference
when the equipment is operated in any environment. This equipment
generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
WARNING Changes or modifications not expressly approved by Welch
Allyn could void the user's authority to operate the equipment.
®
Welch Allyn
Spot
2
b
Pulse modulation
217 Hz
2
b
Pulse modulation
217 Hz
b
0.2
Pulse modulation
217 Hz
Vision Screener Model VS100
0.3
28
0.3
28
0.3
9