Figure 2-6. Flat Surface Mounting Dimensions
2-8
2. The ambient temperature of the transmitter housing must not exceed
65°C (149°F). Locate the electronics in an area where temperature
extremes, vibration, and electromagnetic and radio frequency
interference are minimal.
3. Locate the Rosemount 5081 transmitter within 150 ft (45,7 m) of the
oxygen probe due to wiring and signal considerations.
1. Ensure all components are available to install the Rosemount 5081
transmitter.
2. Choose a method or location to mount the transmitter.
a. Flat Surface Mounting. The transmitter may be mounted on a flat
surface using the threaded mounting holes located on the bottom of
the transmitter housing. Refer to Figure 2-6 for installation
references.
b. Pipe Mounting. An optional pipe mounting bracket is available for
this type of installation. Refer to Figure 2-7 for installation references.
Reference Manual
00809-0100-4882, Rev AB
May 2019