Two-wire in-situ oxygen analyzer (550 c to 1600 c)
Table of Contents

Advertisement

Rosemount 5081FG

Navigation

Figure 3-6. Infrared Remote
Control (IRC)
3-6
The IRC in Figure 3-6 is used to interact with the Rosemount 5081 transmitter
and navigate through the screens on the LCD.
1. Hold the IRC within 6 ft (1,8 m) of the Rosemount 5081 transmitter and
within 15 degrees from the centerline of the transmitter LCD. The
amount of ambient light may also affect IRC performance:
NOTE
The LCD may react slowly to IRC commands. Allow sufficient time between
key presses to avoid undesired or repeated commands from accumulating in
the command queue.
2. Use the keys on the IRC to navigate through the menu screens. Refer to
Figure 3-6. General usage is as follows:
a. RESET. Returns to the PROCESS DISPLAY screen at the top of the
menu tree. Any non-entered number in the exited state will be
ignored, and the previous data will be used.
b. HOLD. Not used.
c. Left/Right Arrow. Moves left and right among editable digits on the
display.
d. Up/Down Arrow. Increases or decreases the value of the currently
selected digit on the display.
e. CAL. Accesses the CALCHECK MENU branch of the menu tree.
Only works from the PROCESS DISPLAY screen.
Reference Manual
00809-0100-4882, Rev AB
May 2019