Oxygen Probe Signal Connections - Emerson Rosemount 5081FG Reference Manual

Two-wire in-situ oxygen analyzer (550 c to 1600 c)
Table of Contents

Advertisement

Rosemount 5081FG
Oxygen Probe Signal
Connections
Figure 2-9. Oxygen Probe
Terminal Block
2-12
Moisture accumulation in the transmitter housing can affect its performance and may void
its warranty.
It is necessary to prevent moisture from entering the Rosemount 5081
transmitter housing. The use of weather-tight cable glands is required. If
conduit is used, plug and seal connections on the transmitter housing to
prevent moisture accumulation in the terminal side of the housing.
1. Two signals represent the O
provides these values to the Rosemount 5081 transmitter for processing
and signal conditioning.
2. Wiring connections for the probe are shown in Figure 2-9.
Reference Manual
00809-0100-4882, Rev AB
value and the cell temperature. The probe
2
May 2019