Installation Specifications - Rosemount 6888 Xi With Rosemount 6888C Transmitter Probe - Emerson Rosemount 6888C Instruction Manual

In-situ combustion oxygen analyzer for use in hazardous areas
Hide thumbs Also See for Rosemount 6888C:
Table of Contents

Advertisement

Specifications
July 2018
3.4.8
Traditional architecture cable
200 ft (61 m) maximum length
3.4.9
Power consumption of transmitter probe/integral
autocal probe
776 W maximum during warm-up
3.5
Installation specifications - Rosemount 6888 Xi
with Rosemount 6888C Transmitter probe
3.5.1
Electrical power of optional Rosemount 6888 Xi
Electronics
120/240 V ±10%, 50/60 Hz
3.5.2
Power consumption of Rosemount 6888 Xi
10 W maximum
3.5.3
Rosemount 6888 Xi alarm relays
2 provided- 2 amps, 20 Vdc
3.5.4
Rosemount 6888 Xi optional loss of flame contact
Removes heater power
3.5.5
Electrical noise
Meets EN 61326, Class A
3.5.6
Traditional architecture cable
200 ft (61 m) maximum length
3.5.7
Transmitter electrical 4-20 mA power
12 - 42 Vdc (looped-powered from the control room or from the Rosemount 6888 Xi box)
18
Instruction Manual
00809-0100-4891
6888C