Reference Manual
00809-0100-4882, Rev AB
May 2019
Reset Max Cell T
Set O
Filter Time
2
Figure 3-12. Reset Max Cell T
Figure 3-13. Set O
Filter Time
2
Refer to Figure 3-12. The transmitter tracks the maximum cell temperature
obtained. Use this screen to reset the maximum cell temperature attained
value to the current cell temperature. Press ENTER to begin editing. Use the
arrow keys to select and change the value (Y/N). Then, press ENTER to
accept the value. Pressing NEXT displays the SET O
Refer to Figure 3-13. In some applications it is beneficial to dampen the raw
O
signal coming from the cell. Use this screen to enter the amount of time it
2
will take the O
to reach 90% of the new reading. Press ENTER to begin
2
editing. Use the arrow keys to select and change the screen value to the O
filter value (in seconds). Enter a value between 0 and 300 seconds and press
ENTER to accept the value. Press NEXT to access the TRIM 4 mA? screen.
Rosemount 5081FG
FILTER TIME screen.
2
2
3-13