Emerson Rosemount 5081FG Reference Manual page 85

Two-wire in-situ oxygen analyzer (550 c to 1600 c)
Table of Contents

Advertisement

Reference Manual
00809-0100-4882, Rev AB
May 2019
Explosion-proof for Class I, Groups B, C, D; Class II, Groups E, F, G, Class III, T6 Tamb = 70 °C
Type 4X
Europe
ATEX
Certificate: Baseefa04ATEX0052X
Standards: EN 60079-0:2012+A11:2013
EN 60079-11:2012
Markings:
II 1 G
Ex ia IIC T4 Ga
(-20°≤ Ta ≤ +65°C)
Special Conditions for Safe Use (X):
1.
The equipment enclosure may contain light metals. The equipment must be installed in such a manner as to minimize the risk of impact or
friction with other metal surfaces
International
IECEx
Certificate: IECEx BAS 09.0159X
Standards: IEC 60079-0:2011
IEC 60079-11:2011
Markings: Ex ia IIC T4 Ga
(-20°≤ Ta ≤ +65°C)
Special Conditions for Safe Use (X):
1.
The Model 5081 enclosure may be made of aluminum alloy and given a protective polyurethane paint finish; however, care should be
taken to protect it from impact or abrasion if located in a zone 0 environment.
Rosemount 5081FG

Advertisement

Table of Contents
loading

Table of Contents