Enabling Posterior Cue & Anterior Cue
1. Select the Cue Tab. A drop-down menu will be displayed. See Figure 6-1.
2. Open the drop-down menu and select the Posterior Cue or Anterior Cue mode. See Figure 6-2.
3. If desired, select the Homing option to require
the trolley to return to an overhead position
when the patient is stationary. See Figure 6-3.
Deselecting the Homing option will maintain an
angled position when the patient is stationary.
Note: The Homing option is enabled by default.
4. Cue Angle may be adjusted to increase or decrease
the rope angle from the vertical orientation by
selecting the Minimum, Moderate, or Maximum
radio buttons. See Figure 6-3.
Note: The trolley will remain in its overhead position before the patient begins walking.
Enabling Anterior Chase
1. Select the Cue Tab from the training screen. A drop-down menu will be displayed. See Figure 6-2.
2. From the drop-down menu, select Anterior Chase.
3. The patient body weight will be displayed in the Cue Tab, and this value may be adjusted from the
Training Screen as needed to ensure accuracy.
4. Set the target speed by pressing the plus (+) and minus (-) buttons. See Figure 6-4.
32
Clinician's Guide
Figure 6-1: Advanced Trolley Tracking Cue Tab
Figure 6-3:
Homing & Cue Angle
Figure 6-2:
Trolley Cue Mode
Drop-Down Menu
Figure 6-4:
Anterior Chase
Controls
Need help?
Do you have a question about the Vector Elite and is the answer not in the manual?