Alignment of z-drive

Advertisement

14. Turn on the autosampler. It should home x, y, and z-axis.
15. Now open the program that you downloaded off the website. It will open a tcomm debug window,
then soon after will open the main window as shown in figure 1-5.
Figure 1-5
16. In the top left hand corner of the main window it should tell you what version of e-prom is in your
autosampler.
17. Click the Home button and unit should act just as it did when you turned it on.
18. The # next to Tray is the # of tubes in a rack, please enter that # now.
19. The number next to Down is the depth that the Z-drive travels. Click the arrow till the number reads
150 for a normal z-drive and 100 for a nicolet z-drive.
20. Now under the Exercise window click Corners.
21. When the autosampler gets to the first tube and z-drive is down turn autosampler off. Align the sample
probe (the one on the right side) so it is in the middle of the test tube.as shown in figure 1-6. Tighten
the cap screws slightly on that side.
Figure 1-6