Alignment of z-drive

Advertisement

22. Turn autosampler back on and when the sample probe gets to the next tube, do the same as you did on
the first tube. Turn unit off align and tighten screw. As shown in figure 1-7.
Figure 1-7
23. Turn unit back on and let unit run through all the 4 corners.
24. If the alignment is off slightly when it hits the 2 tubes in the front of the tray, then split the difference
between the back tubes and the front tubes.
25. If satisfied with the alignment, then tighten all 4 cap screws as tight as possible. The alignment may
move when tightening the screws so check before closing software.
26. Click the close button on the software before turning off the autosampler for the final time, or you may
get an error message.
27. At this time the autosampler should be aligned and ready for your application.