78 Functional verification and calibration overview
Max: 5 mmHg
Mode
Adult mode:
Target pressure ± 5
mmHg
0 mmHg
50 mmHg
150 mmHg
250 mmHg
SpO2 tests
Test
SpO2 pulse rate
60 bpm
SpO2 saturation
90%
Thermometry tests
Temperature test
97.3 °F (36.3°C)
Temperature
tested
96.8 °F (36.0 °C)
101.3 °F (38.5 °C)
105.8 °F (41.0 °C)
Specification
280 . . . . 329 mmHg
Pressure meter
mmHg
mmHg
mmHg
mmHg
Specification
Pulse rate 60 ±1 bpm
59 to 61 bpm
Saturation 90% ±1 sat
point
89 to 91%
Specification ± 0.2°F (± 0.1°C)
97.1 to 97.5°F
36.2 to 36.4°C
Specification ± 0.2 °F (± 0.1 °C)
96.6 °F to 97.0 °F
35.9 °C to 36.1 °C
101.1 °F to 101.5 °F
38.4 °C to 38.6 °C
105.6 °F to 106.0 °F
40.9 °C to 41.1 °C
mmHg
Actual reading
Specificati
Service tool
on
mmHg ± 1 mmHg
mmHg ± 3 mmHg
mmHg ± 3 mmHg
mmHg ± 3 mmHg
Actual reading
bpm
%
Actual reading
°F or °C
Actual reading
°F or °C
°F or °C
°F or °C
mmHg
Pass
Fail
Pass
Fail
Pass
Fail
Pass
Fail