Message
Temperature not functional. 03CB00 Unable to send message
Temperature not functional.
03CC00
Temperature not functional.
03CD00
Temperature not functional. 03CE00 Unable to read PIM file
Temperature not functional. 03CE01 Upgrade file directory
Direct mode reading timed out
Tissue contact lost.
Temperature module reset. 03D000 Temperature sensor reset
Patient data messages
Message
Database schema out of
data; recreating.
Database is unreadable
during startup;
recreating. 1F0001
Error accessing PDM
database; restarting PDM.
1F0002
Maximum number of
patient records + Oldest
record overwritten.
No data saved.
Save successful.
Possible cause
to the Temperature
module
Temperature module
communication times
out
Failed to upgrade the
temperature module
could not be accessed
Direct mode reading
times out
Lost tissue contact while
attempting to acquire a
temperature
measurement or
acquired measurement
was performed with
limited tissue contact.
unexpectedly
Possible cause
The database was cleared due to a
schema update
The database was unreadable
during startup
Database corrupted when device is
in operation
Data was deleted as it contained
more than 50 records
A manual save is not allowed
A manual record was saved
Suggested action
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Retry the device update.
Retry the device update
Direction mode reading times out
Check the tissue contact and try the
measurement again.
None
Suggested action
Information status message; press OK button
to dismiss.
Press OK button to dismiss.
Press OK button to dismiss.
Information status message; press OK button
to dismiss.
Information status message; press OK button
to dismiss.
Information status message; press OK button
to dismiss.