NIBP not functional. 050613
NIBP not functional. 050614
NIBP not functional. 050615
NIBP not functional. 050616
NIBP not functional. 050617
NIBP not functional. 050618
Check cuff inflation settings.
NIBP not functional. 05FF01
NIBP not functional. 05FF02
NIBP not functional. 05FF03
NIBP not functional. 05FF04
NIBP not functional. 05FF05
NIBP not functional. 05FF06
NIBP not functional. 05FF07
Possible cause
NIBP response header wrong
NIBP response checksum wrong
Too much NIBP data was received
NIBP FPROM erase error
NIBP FPROM programming error
Invalid NIBP target pressure
Cuff inflation target was overridden
due to max pressure being too low
Unrecognized WACP parameter
received from sensor
Timeout waiting for sensor
response
Error deserializing the WACP
message received from the sensor
WACP stack send message failure
Timeout waiting for asynchronous
sensor message
One or more undetermined
numerics when reading status
indicates OK
Unrecognized sensor reading status
code
Appendix 109
Suggested action
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Change cuff inflation target or
maximum pressure so that the cuff
inflation target is at least 20 mmHg
lower than maximum pressure.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Check connections. Restrict patient
movement.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.