Welch Allyn Spot Vital Signs 4400 Service Manual page 112

Hide thumbs Also See for Spot Vital Signs 4400:
Table of Contents

Advertisement

108 Appendix
Message
NIBP not functional. 050607
NIBP not functional. 050608
Stat mode requested too soon.
Clear to retry. 050609
Unable to determine NIBP; check
connections and tubing for kinks.
05060A
NIBP not calibrated. Calibrate the
module. 05060B
NIBP not functional. 05060C
NIBP not functional. 05060D
NIBP not functional. 05060E
NIBP not functional. 05060F
NIBP not functional. 050610
NIBP not functional. 050611
NIBP not functional. 050612
Possible cause
NIBP cannot clear the failsafe errors Internal malfunction. If the problem
NIBP safety processor has stopped
responding
NIBP excessive stat mode time. The
time between readings is less than
one minute and the readings plus
time between readings result in the
device taking over 15 minutes to
complete the averaging cycle.
NIBP transducers are not matched
Factory EEPROM checksum error on
NIBP. Units internal configuration
was corrupted
NIBP command not implemented
NIBP wrong data count
NIBP data range error
NIBP no POST error to clear
NIBP cannot clear this POST error
NIBP command not command type Internal malfunction. If the problem
NIBP communication timeout
Suggested action
inspected by a qualified service
technician.
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
The transducer are above 5 mmHg
and the difference in pressure is
greater than 40 mmHg. Check cuff
for pinched or obstructed tubing. If
the problem persists, have the
device inspected by a qualified
service technician.
Calibrate the NIBP module. If the
problem persists, have the device
inspected by a qualified service
technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
persists, have the device inspected
by a qualified service technician.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician..