Self-Diagnostic Warning Light - Invacare Scorpius Series Service Manual

Table of Contents

Advertisement

Scorpius

1.6 Self-Diagnostic Warning Light

When the main power switch is turned on, the self-diagnostic warning light will flash if there be any
malfunction or failure.
A u t o m a t i c V e r s i o n : s e l f - d i a g n o s t i c w a r n i n g l i g h t ,
s h a r e d w i t h t h e b a t t e r y i n d i c a t o r .
N o r m a l :
r e m a i n s o n , a s d o e s t h e b a t t e r y p o w e r
i n d i c a t o r .
F a i l u r e :
f l a s h e s a c c o r d i n g t o t h e n a t u r e o f t h e
f a i l u r e , s e e t h e t a b l e b e l o w .
Number of
Possible Cause
Flashes
1
Battery Low
Low Battery Fault
2
High Battery Fault
3
4
Current limit time-out or controller
overheat
5
N-D lever not set to D-position
Electromagnetic brake failure
The throttle is not in Neutral or the
6
folding/ unfolding procedure is
uncompleted.
The folding/ unfolding procedure is
uncompleted
Scooter battery is being charged
Micro switch failure
7
Speed pot
Motor or relevant circuit failure
8
9
Main cable, controller failure or
controller temperature too high
1654773~A
M a n u a l V e r s i o n : s e l f - d i a g n o s t i c w a r n i n g l i g h t ,
s h a r e d w i t h t h e m a i n p o w e r l i g h t .
N o r m a l :
c o n s t a n t g r e e n l i g h t .
F a i l u r e :
f l a s h e s a c c o r d i n g t o t h e n a t u r e o f f a i l u r e ,
s e e t h e t a b l e b e l o w .
Handling Method
The batteries are running low.
Recharge the batteries.
The batteries have run out of charge.
Recharge the batteries.
Check the battery and associated connections
and wiring.
Handling method:
1. Disconnect the battery pack and check the
battery condition.
2. SERVICE-6 Battery Pack Repair.
Do not drive up steep slopes or overload the scooter.
Switch to D (drive) position. Turn off the power and
turn on again.
SERVICE-13-5 Electromagnetic Brake Repair.
Solution:
Repeat the folding or unfolding procedure.
SERVICE-1-3 Starter VR Repair.
Repeat the folding or unfolding procedure.
Turn off the main power and disconnect the charger.
SERVICE-12-1 Micro Switch Repair.
SERVICE-1 Steering Shroud Repair.
Solution:
SERVICE-11-1 DR Controller Repair
SERVICE-13-4 Motor Repair
Solution:
SERVICE-11-1 DR Controller Repair
SERVICE-2-1 Main Cable Repair
Service Manual
9

Advertisement

Table of Contents
loading

This manual is also suitable for:

ScorpiusScorpius-a

Table of Contents