SERVICE-06-1 Battery Pack - Charger Socket Repair

Please refer to SERVICE-06 Battery Pack Repair steps and then:
1. Find the charger socket connector inside the battery pack (Figure 1 below).
2. Reconnect the battery pack charger socket connector.
3. If this does not restore function, refer to the following Replacement Instructions.
Replacement Instructions:
After completing the SERVICE-06 Battery Pack Repair steps:
1. Remove the battery pack charger socket.
2. Remove the battery pack outer shroud and the charger socket screws.
3. Remove the battery pack charger socket and remove the inside metal sheet at the same time.
4. Take out the battery pack charger socket, and replace it with a new one.
❖ After the replacement of parts or re-installation, please follow the exact reverse sequence for reassembly.
2.6.2

SERVICE-06-2 Battery Pack - Power Socket Repair

Please refer to SERVICE-06 Battery Pack Repair steps and then:
1. Find the battery pack charger connector (Figure 1 below).
2. Reconnect the battery pack charger socket connector.
3. If it is not effective, please refer to the following Replacement Instructions.
Replacement Instructions:
After completing the SERVICE-06 Battery Pack Repair steps:
1. Remove the battery pack charger socket.
2. Squeeze the two sides of the charger socket inside the battery pack.
3. Push it out of the battery pack gently.
4. Replace the battery pack power socket with a new one.